Obtaining meaningful data can have a significant impact on your process whether using the mercury CEMS for process monitoring or compliance monitoring. Accurate and relevant data can result in cost and resource savings. The table below lists some of the required tests and recommended on- going checks using the Mercury CEMS for either compliance or process control in order to provide the greatest amount of meaningful data.
Note that the acceptance criteria for compliance monitoring falls under the MATS or the PC MACT rule and this is not a complete list. For process control applications, the pass/fail criteria can be user-specific depending on the application.
System Check / Configuration | Compliance Monitoring | Process Control |
---|---|---|
Zero & Span | Required Daily | Strongly Recommended Daily |
System Integrity Test | Required Weekly | Optional |
Linearity | Required Quarterly | Recommended |
Elemental Generator Audit | Required Monthly | Recommended Semi-Annually |
RATA | Required Annual | Partial RATA Recommended |
Speciation Mode | Total Hg Only Required | Speciation Recommended |
7 Day Drift Test | Required for Initial Certification | Recommended upon install |
Cycle response time | Required for Initial Certification | Recommended upon install |
Probe Type | Multiple Options | Inertial Probe Recommended for dirty applications |