Perform ion analysis with the Thermo Scientific Dionex Inuvion IC System, an intuitive, ultra-reliable ion chromatography (IC) instrument with straightforward operation and extended application capabilities. This system supports electrolytic suppression and can include Reagent-Free IC (RFIC) eluent generation for additional time-savings, greater day-to-day consistency, and gradients to optimize methods. It offers automated start-up and shut-down routines to save time and consistently reliable performance for busy analytical testing laboratories.
Flexible, adaptable platform
The Dionex Inuvion IC System is designed to help your lab meet the needs of today and the future. Upgrade the Dionex Inuvion IC System to include RFIC for automated eluent generation capabilities. Optional accessories can be swapped in and out of the instrument by users in less than five minutes for extended application capabilities.
Ultra-reliable performance
This state-of-the-art ion chromatography system is engineered for maximum productivity. With new pump technology and electronics, the Dionex Inuvion IC System provides enhanced day-to-day reliability. System self-diagnostics also detect any issues with hardware and consumables, giving you the confidence that it will be ready to run whenever you are.
Easy-to-use platform
The Dionex Inuvion IC System is designed for fast startup and maximum uptime. With convenient, unobstructed access to all necessary components and a logical flow path, users can quickly and safely access everything on the instrument.
Suppression compatibility
The Dionex Inuvion IC System offers the choice of using either chemical suppression, for more challenging samples, or electrolytic suppression, for ease of use and convenience. Suppressors remove conductive ions from the eluent to decrease background conductivity, while enhancing analyte signal, enabling high sensitivity to achieve low detection limits with conductivity detection.
High-pressure operation
With the ability to operate at up to 5,000 psi, the Dionex Inuvion IC System can take advantage of high-performance 4 µm-particle-size Thermo Scientific Dionex IC columns for more efficient separations that can be used to shorten run times.
Chromatography data system
The Dionex Inuvion IC System comes with the easy-to-use Thermo Scientific Chromeleon Chromatography Data System (CDS). This CDS platform automates and optimizes daily tasks and helps even inexperienced users achieve consistent, high-quality results.
The table below provides an overview of the Dionex Inuvion IC System specifications. Not all specifications are listed. For a full list of the specifications, download the Dionex Inuvion IC System specifications sheet.
Specification |
Dionex Inuvion IC System |
Analytical pump and fluidics |
|
Type |
Serial dual-reciprocating pistons, microprocessor-controlled constant stroke, variable speed |
Construction |
Chemically inert, metal-free PEEK pump heads and flow paths compatible with aqueous eluents of pH 0–14 and reversed-phase solvents |
Pump operating pressure |
0–35 MPa (0–5000 psi) |
Flow rate range |
0.00–5.00 mL/min in 0.01 mL/min increments |
Flow precision |
<0.1%, typically |
Flow accuracy |
<0.1%, typically |
Pressure ripple |
<1% |
Eluent on-off valve |
Standard |
Leak sensor |
Optical, standard |
Piston seal wash (optional) |
Pump head wash can be operated in continuous or intermittent mode when connected to rinse solution supply |
Pressure alarm limits |
Upper and lower limit pressure alarms can be set |
Vacuum degas |
Standard, user adjustable vacuum level |
Eluent bottles |
Standard 2 L polypropylene bottle; allows various sizes |
Eluent bottle pressure regulator (optional) |
Supported by digital controlled regulator with display |
Injection valve |
6-port, 2-position Rheodyne valve, electronically activated |
Columns supported |
2, 3, 4, and 5 mm ID; maximum length 250 mm analytical column with 50 mm guard column |
Eluent generator (RFIC configuration) |
|
Eluent types |
KOH, MSA, or K2CO3 |
Eluent concentration range |
0.1-100 mM |
Flow rates |
0.10–3.00 mL/min when a Thermo Scientific Dionex EGC is installed |
Maximum operating pressure |
Dionex EGC cartridges: 35 MPa (5000 psi) |
Gradient profiles |
Up to 9 time-defined gradient steps |
Column heater (optional) |
|
Operating temperature range |
10 to 60 °C (50 to 140 °F); settable within the software; minimum working range is 5 °C above ambient temperature |
Temperature accuracy |
±0.5 °C at sensor, at calibration points (35, 45 °C) |
Suppressors and control |
|
Chemical and electrolytic suppression |
2 mm and 4 mm anion and cation suppressor types |
Suppressor regeneration mode |
Electrolytic suppressor – Recycle or external water mode Chemical suppressor – External regenerant |
Current control range |
Thermo Scientific Dionex DRS 600 Electrolytically Regenerated Suppressor: 0–500 mA (4 mm) and 0–150 mA (2 mm) in 1 mA increments Thermo Scientific Dionex ERD 500 Electrolytically Regenerated Desalter: 0–500 mA (4 mm) and 0–150 mA (2 mm) in 1 mA increments |
Salt converter |
Thermo Scientific Dionex SC-CERS 500 available in 2 and 4 mm versions |
Carbonic acid removal for anions |
Thermo Scientific Dionex CRD 200 Carbonate Removal Device for use with hydroxide eluents or Dionex CRD 300 for use with carbonate eluents |
Non-suppressed chromatography |
Yes, supported |
Suppressor wear parts |
Chemical suppression: optional regenerant pump Electrolytic suppression: none |
Dynamic suppression capacity |
Anions:
Cations:
|
Void volumes |
|
Conductivity detector electronics and flow cell |
|
Type |
Microprocessor-controlled digital signal processor |
Cell drive |
128 kHz square wave |
Linearity |
r2≥0.999% |
Resolution |
0.002 nS/cm |
Full-scale output ranges |
Digital input signal range 0-18,000 μS/cm, with auto-ranging; analog output signal range 0-18,000 μS/cm |
Temperature compensation |
Variable, default set at 1.7%/°C at cell temperature |
Temperature range |
Ambient +7 °C, 15 to 60 °C |
Cell electrodes |
Passivated 316 stainless steel; compatible with MSA |
Cell body |
Chemically inert polymeric material |
Cell volume |
<1 µL |
Heat exchanger |
Inert, tortuous path for low axial dispersion |
Maximum cell operating pressure |
10 MPa (1,500 psi) |
Data filter |
Rise times from 0 to 10 s, Data Collection Rate 1 to 100 Hz, user selectable |
Autosampler |
|
Automation using autosampler |
For more information about available autosamplers, visit IC Autosamplers |
System software |
|
|
Chromeleon CDS, requires Microsoft® Windows® 10 or 11 |
Physical specifications |
|
Dimensions (h × w × d) |
66.1 x 29.2 x 43.2 cm (26.0 x 11.5 x 17.0 in.) |
Weight |
16.2 kg (36 lb) |
Power requirements |
100–240 V AC, 50–60 Hz autoranging |
Operating temperature |
4–40 °C (40–104 °F) |
Operating humidity range |
20-80% relative, non-condensing |
Control modes |
Full control through Chromeleon CDS software; alternative control through TTL or relay closures; one relay output, two TTL outputs, two assignable TTL inputs |
USB communication protocol |
One USB input; three USB outputs |
For Research Use Only. Not for use in diagnostic procedures.