Vanquish™ Dual Split Sampler
Vanquish™ Dual Split Sampler
Actual product may vary
Thermo Scientific™

Vanquish™ Dual Split Sampler

Green features

Improve productivity with two injection devices in one housing, resulting in two separate channels that are individually controllable. Available in 1000 bar or 1500 bar options.

Have Questions?
Change viewbuttonViewtableView
Catalog NumberDescription
VF-A40-A-02Vanquish Dual Split Sampler FT
VH-A40-A-02Vanquish Dual Split Sampler HT
Catalog number VF-A40-A-02
Price (USD)/ Each
-
Request A Quote
Description:
Vanquish Dual Split Sampler FT

Improve productivity using the Thermo Scientific™ Vanquish™ Dual Split Sampler. Designed to support Thermo Scientific™ Vanquish™ Duo UHPLC system for Dual LC, this unique biocompatible autosampler combines two injection devices in one housing, resulting in two separate channels that are individually controllable (injector volume, wash cycles, etc.). Choose between the 1000 bar or 1500 bar option.

View HPLC/UHPLC resources

Dual sample injection—highest productivity, fastest return on investment

Vanquish Dual Split Samplers are made for Vanquish Duo system for Dual LC. These fully biocompatible autosamplers are optimized for the highest level of ruggedness and uptime even under tough system pressure and eluent conditions.

  • Improves productivity by using two independent injections units controlled independently from a single user interface with full traceability with Thermo Scientific™ Chromeleon™ Chromatography Data System (CDS)
  • Increases sample rack throughput and capacity without changing validated methods
  • Increase capacity without sacrificing bench space
  • Tracks sample configuration with automated barcode reading
  • Boosts return on investment by reducing cost per sample
  • Analyzes the same set of samples with multiple methods reducing sample preparation time and maximizing sample information
Specifications
TypeDual Split Sampler FT
Injection Cycle TimeDown to 8 s depending on separation conditions
Injection Volume Precision< 0.25% area RSD for 1 μL (caffeine aq. soln.)<br /> Typically < 0.5% area RSD for 0.5 μL (caffeine aq. soln.)
Injection Volume Range0.01 to 25 μL, min. step = 0.01 μL; Optional: 0.01 to 100 μL
Injection TypeSplit Loop, 2 Injection Units
Sample Capacity4 Segments for Sample Racks or Well Plates with SBS Footprint Format; 12 x 22.5 mm O.D. Vials
Max. Pressure1034 bar
Biocompatible VersionYes, pH range 2–12, chloride concentration up to 1mol/L
Wetted Parts: Sample flow path: Titanium, Ceramics, PEEK, MP35N, DLC
Eluent Flow Path: MP35N, Titanium, Sapphire, PEEK, PTFE, Ceramics, DLC
Wash Liquid Flow Path: Silicone, PP, FFPM, PEEK, PA
PC ConnectionUSB 2.0; 3-port-HUB to connect further Vanquish modules
Weight (Metric)29 kg
DescriptionVanquish Dual Split Sampler FT
Carryover< 0.002% with caffeine (typically: < 0.0004%), UV
Unit SizeEach