Specifications
Flow Rate1 L/min.
Inputs10 Digital Inputs (Standard), 16 Digital Inputs
Linearity±2% full scale
Response Time70 sec. (except for oxygen (electrochemical cell: 60 sec.; paramagnetic cell: 45 sec.)
Temperature (English) Operating41°F to 113°F in non-condensing environment
Temperature (Metric) Operating5°C to 45°C
Certifications/ComplianceDesigned to meet US EPA 40CFR Part 60 requirements
Depth (English)23 in.
Depth (Metric)584 mm
DescriptionNDIR Multi-Gas Analyzer
Height (English)8.62 in.
Height (Metric)21.9 cm
Measurement RangeO: 0 to 5% (min. range), 0 to 25% (extended range), 0.01% (lower detection limit); CO: 0 to 100ppm (min. range), 0 to 2500ppm (extended range), 1ppm (lower detection limit); CO2: 0 to 5% (min. range), 0 to 25% (extended range), 500ppm (lower detection limit)NO: 0 to 100ppm (min. range), 0 to 2000ppm (extended range), 1ppm (lower detection limit)NO2: 0 to 50ppm (min. range), 0 to 500ppm (extended range), 0.2ppm (lower detection limit)SO2: 0 to 100ppm (min. range), 0 to 10,000ppm (extended range), 0.2ppm (lower detection limit)
OutputsSix analog outputs selectable voltage; six additional optional outputs are availale
Voltage100/150/220/240 V
Weight (English)49 lb.
Weight (Metric)22.2 kg
Width (English)16.75 in.
Width (Metric)425 mm
Zero DriftO [Electrochemical Cell]: Zero Drift (24 hr./7 day), <0.2%/<0.2%; Span Drift (24 hr./7 day), <0.2%/<0.3%; O [Paramagnetic Cell]: Zero Drift (24 hr./7 day), <0.1%/<0.1%; Span Drift (24 hr./7 day), <0.1%/<0.2%; CO: Zero Drift (24 hr./7 day), <1ppm/<3ppm; Span Drift (24 hr./7 day), <1% of range/<3% of range.CO2: Zero Drift (24 hr./7 day), <0.1%/<0.5%; Span Drift (24 hr./7 day), <1% of range/<3% of range.NO: Zero Drift (24 hr./7 day), <1ppm/<3ppm; Span Drift (24 hr./7 day), <1% of range/<3% of range.NO2: Zero Drift (24 hr./7 day), <1ppm/<3ppm; Span Drift (24 hr./7 day), <1% of range/<3% of range.SO2: Zero Drift (24 hr./7 day), <1ppm/<3ppm; Span Drift (24 hr./7 day), <1% of range/<3% of range.
Unit SizeEach