Checking Your T4 DNA Ligase Activity

Troubles with ligation?

Check the activity of your T4 DNA Ligase with two easy-to-do control experiments

You may check the efficiency of your ligation reaction by mixing the reaction with loading dye containing SDS (final SDS concentration 0,2%) and running it on a standard agarose gel. After ligation several high molecular weight bands should appear. Without SDS in the loading dye a smear will appear and it will be difficult to judge the ligation result. (Figure 1).

If the DNA is not ligated you may want to check the activity of your ligase by using lambda DNA digested with HindIII (Cat. No. SM0101) as substrate DNA. This DNA is highly pure and should be easily ligated using the protocol below.


Protocol

  • Lambda DNA/HindIII
  • 2 µL 10x T4 DNA Ligase buffer
  • Fill with water to 19 µL
  • Add 1 Unit T4 DNA Ligase (you may dilute it in 1x T4 DNA Ligase Buffer)

Total volume: 20 µL
Incubate for 10 min at 22°C (room temperature)

Analyse 10 µL of this reaction in a gel as described using a loading dye with SDS (final SDS concentration 0.2%). After ligation you will see several high molecular weight bands as shown in Figure 2.

If the ligation works perfectly fine for Lambda/ HindIII but not for your substrate DNA the ligase is active but the substrate is not suitable for ligation. This may be due to several reasons and our Technical Support will be happy to assist you with troubleshooting.

Figure 1

Figure 1

Lane 1: before ligation

Lane 2: after ligation, loading dye without SDS

Lane 3: after ligation, loading dye with SDS 

Lane 4: Thermo Scientific GeneRuler DNA Ladder Mix (Cat. No. SM0331)

Figure 2

Figure 2

Lane 1: Lambda/Hindlll before ligation

Lane 2: Lambda/Hindlll after ligation

仅供科研使用,不可用于诊断目的。